ECD Introduces Free Chlorine Analyzer for Produce Washing, Food Processing Applications

Electro-Chemical Devices (ECD) has introduced its FC80 High Range Free Chlorine Analyzer, designed to monitor free chlorine concentrations in water used for produce washing, food processing sanitation, and disinfection applications. The analyzer measures free chlorine concentrations ranging from 0.05–200 parts per million (ppm) and features automatic pH compensation and built-in flow control.
The FC80 is intended for applications including the disinfection of produce wash water, cleaning and disinfection of water supply lines, and monitoring of cleaning solutions. Chlorine-based disinfection is used in food processing operations and produce packing lines to help control microbial contamination, including pathogens such as Escherichia coli and Salmonella.
According to the U.S. Food and Drug Administration (FDA), liquid chlorine and hypochlorites are generally used at concentrations of 50–200 ppm, with contact times of 1–2 minutes, to sanitize produce surfaces and processing equipment.
Hypochlorous acid (HOCl) is the form of free chlorine with the highest bactericidal activity against a broad range of microorganisms. In water-based solutions, the equilibrium between HOCl and hypochlorite ions (OCl⁻) is pH-dependent, with HOCl concentrations increasing as pH decreases. Other factors affecting HOCl concentrations include temperature, organic matter, light, air, and metals.
The FC80 features a panel-mounted design incorporating a constant-head flow control device, pH sensor, chlorine sensor, and ECD's T80 analyzer/transmitter, all mounted on a polyvinyl chloride (PVC) panel.
Built-in flow control eliminates the need for separate pressure regulators and rotameters, while automatic pH compensation eliminates the need for chemical reagents used in sample conditioning.
The analyzer's free chlorine sensor measures the HOCl component of free chlorine, while the transmitter calculates the remaining free chlorine concentration using either measured pH or a user-defined fixed value. The integrated pH sensor provides compensation for samples with pH values between 6 and 9.
Looking for quick answers on food safety topics?
Try Ask FSM, our new smart AI search tool.
Ask FSM →
The FC80 can be installed by connecting the sample and drain lines, outputs, and power supply. Calibration and standardization are performed through comparison with the N,N-diethyl-p-phenylenediamine (DPD) method.
The analyzer is available with either 110–240 VAC or 24 VDC power. Measurement parameters can be displayed using user-defined line, bar, or gauge graphs.
The standard configuration includes two 4–20 mA outputs, Modbus RTU communication, and three single-pole, double-throw (SPDT) relays. The relays can be configured for high/low alarms with expiration timers, periodic timers, or fault alarms.
An optional automatic cleaning feature includes a solenoid-actuated spray cleaner that uses either process water at 30 pounds per square inch (psi) or compressed air. An adjustable timer controls the frequency and duration of cleaning cycles.
ECD: www.ecdi.com